855M495 1.0INNOVATIVE PC Data Acquisition SolutionsUDAS-1001E SeriesHardware User Manualfor theMultifunction Data Acquisition System

Pagina 2

1- 3UDAS SYSCHECKThe SYSCHECK application enables you to verify the UDAS installation is functioning properly. SYSCHECK can also be used during the ca

Pagina 3

UDAS-1001E Series Data Acquisition Systems1-4Digital Input/OutputDigital Input/Output• 16 Programmable digital I/O channels. Groups of 8 separately pr

Pagina 4 - TrademarksTrademarks

1- 5• Programmable gains are 1, 10, and 100 which translate to a full-scale input of 0-10 V, 0-1.0 V, 0-0.1 V, ±10 V, ±1.0 V, ±0.1 V, ±5 V, ±0.5 V, a

Pagina 5 - 1.0 000218 Original Release

UDAS-1001E Series Data Acquisition Systems1-6If an analog input is used for the trigger, specify the trigger level, analog input channel, and slope. U

Pagina 6 - Table of Contents

1- 7The scanner operates with the built-in analog channels, or with the optional AI-MAX. The AI-MAX allows up to 256 analog input channels to be conne

Pagina 7 - Appendix B Accessories

UDAS-1001E Series Data Acquisition Systems1-8Example:If 9 channels are being scanned at a rate of 100Hz and the slope is low to high, the trigger sign

Pagina 8

1- 9Digital I/O signals are TTL-compatible and perform functions such as reading digital inputs, driving digital outputs, and detecting input change-o

Pagina 9 - 1.1.11.1.1 FeaturesFeatures

UDAS-1001E Series Data Acquisition Systems1-10A diagram of the termination panel for the UDAS-1001E-4 unit is shown below.FIGURE 1.2 UDAS-1001E-4 dime

Pagina 10 - Analog OutputAnalog Output

1- 111.2.21.2.2 PowerPowerThe UDAS unit normally uses power from the host computer via the USB cable. Power can also be obtained from an external powe

Pagina 11 - TerminationTermination

UDAS-1001E Series Data Acquisition Systems1-12Visual Designer UDAS Support LibraryThis support package provides the block functions necessary to acqui

Pagina 12

Copyright 2000 by Intelligent Instrumentation Inc., Tucson, Arizona, USAAll rights reserved.Warranty and Repair Policy Statement Warranty and Repair P

Pagina 13 - Rate GeneratorRate Generator

2-1Chapter 2Chapter 2 Configuration and Configuration and InstallationInstallation2.12.1 IntroductionIntroductionThis chapter contains instructions on

Pagina 14

Configuration and Installation2-2Setting the ConfigurationSetting the ConfigurationSwitches 1-8 set the initial levels of the port’s channels 0-7 if t

Pagina 15 - Digital I/ODigital I/O

2-3If you plan to use any of the digital I/O ports for output control in your application(s), note the following Caution statement:Locate switch S1 sh

Pagina 16 - CountersCounters

Configuration and Installation2-4Channels connected to a thermocouple must be connected to bias resistors for proper operation. Use jumpers JP2 and JP

Pagina 17

2-52.32.3 Connecting the UDAS UnitConnecting the UDAS UnitMaking the Initial ConnectionMaking the Initial Connection1. With the computer powered up, c

Pagina 18 - 1.2.21.2.2 PowerPower

3-1Chapter 3Chapter 3 Checking the SystemChecking the System3.13.1 IntroductionIntroductionThis chapter contains instructions for installing and using

Pagina 19

Checking the System3-23.23.2 Installing and Running UDAS SYSCHECKInstalling and Running UDAS SYSCHECKUDAS SYSCHECK provides an easy-to-use graphical i

Pagina 20 - InstallationInstallation

3-33.3.1 Unit InfoThe Unit Info dialog displays the number of UDAS units detected and has a drop down list containing the serial number for each UDAS

Pagina 21 - FIGURE 2.1 UDAS Switch S1

Checking the System3-43.3.2 System CheckWhen the System Check tab is selected, a dialog containing buttons used to execute a variety of system test pr

Pagina 22 - CAUTION:

3-53.3.2.13.3.2.1 Analog Input TestAnalog Input TestClick the Analog In button to access a new window specific to analog input. The UDAS SYSCHECK ANAL

Pagina 23

Technical Assistance and ServiceTechnical Assistance and ServiceSeller's warranty as herein set forth shall not be enlarged, diminished or affect

Pagina 24

Checking the System3-63.3.2.23.3.2.2 Analog Output TestAnalog Output TestThe analog output test controls the voltage on the analog output channels of

Pagina 25 - WARNING!

3-73.3.2.33.3.2.3 Counter TestCounter TestThe counter test monitors TTL transitions that are connected to the counter input of the UDAS unit. Dependin

Pagina 26 - Checking the System

Checking the System3-8For the Count when gate is high and the Count when gate is low modes, the TTL level of the gate enables or disables counting, de

Pagina 27 - 3.3.1 Unit Info

3-9Select the desired Gate Time. The data displays shown at the lower portion of the dialog provide the status and the frequency in hertz.FIGURE 3.7 U

Pagina 28 - 3.3.2 System Check

Checking the System3-10Digital I/O TestDigital I/O TestThe Digital Input/Output test enables you to read data from, or write data to, either 8-bit Dig

Pagina 29

3-113.3.3 Paced Analog InputThe Paced Analog Input dialog enables you to configure a hardware-paced analog input process.Select the number of channels

Pagina 30

Checking the System3-12Clicking the High-Speed Run button (shown in FIGURE 3.9, UDAS SYSCHECK Paced Analog Input Window) configures and starts the pro

Pagina 31

3-133.43.4 Troubleshooting TipsTroubleshooting Tips3.4.1 Basic TroubleshootingIf you are unable to successfully communicate with the UDAS unit, this s

Pagina 32

Checking the System3-14UDAS DriversUDAS DriversVerify the UDAS system driver is properly installed by following the instructions for the USB Port (abo

Pagina 33

3-15• Check the wiring and cables for shorts or damage.If errors still persist after repeatedly checking the unit, check the technical support links o

Pagina 34

TrademarksTrademarksU-LinkSoftware Libraries is a trademark of Intelligent Instrumentation Incorporated.VisualDesigner is a registered trademark

Pagina 35 - 3.3.3 Paced Analog Input

A-1Appendix AAppendix A SpecificationsSpecificationsA.1A.1 UDAS-1001E Series Hardware SpecificationsUDAS-1001E Series Hardware SpecificationsAll speci

Pagina 36

SpecificationsA-2Noise Input grounded at connector (RMS)Gain <=10 1.0 LSBGain = 100 1.2DC Accuracy No missing codes, guaranteedLinearity Error ±0.0

Pagina 37 - 3.4.1 Basic Troubleshooting

A-3Resolution 32-BitNumber of Channels 1Max Input frequency TTL-level input 2 MHzTimeBase GeneratorType Crystal-based Rate Generator (24-Bit)Number 2

Pagina 38 - Power TestPower Test

SpecificationsA-4A.2A.2 UDAS-1001E-3, UDAS-1001E-4 DimensionsUDAS-1001E-3, UDAS-1001E-4 DimensionsUDAS units are housed in standard metal enclosures m

Pagina 39

A-5FIGURE A.3 Dimensions of the UDAS unit (Bottom view)4-40 FH SCREW1.295.5254.7004 PL4.700

Pagina 40 - SpecificationsSpecifications

SpecificationsA-6The dimensions of the UDAS PC board are shown below in FIGURE A.4 Dimensional Drawing of the UDAS series PCB.FIGURE A.4 Dimensional D

Pagina 41 - Specifications

A-7A.3A.3 UDAS I/O ConnectorsUDAS I/O ConnectorsSignal Name Pin Single Ended Input Function Differential Input FunctionAI0 1 Analog Input Channel 0 An

Pagina 42 - (approximately)

SpecificationsA-8AO 0 21 Analog Output Channel 0 **AO 1 21 Digital I/O bit **Digital Port 0 bit 0 27 Digital I/O bit **Digital Port 0 bit 1 28 Digital

Pagina 43

A-9External Input 47 External Trigger or Pacer input **Digital Ground 49 Ground **Digital Ground 50 Ground **Unused 12 **Unused 45 **Signal Name Pin

Pagina 44

B-1Appendix BAppendix B AccessoriesAccessoriesThis appendix includes information on the termination and mounting accessories that are available separa

Pagina 45

UDAS-1001E Series User Manual Revision HistoryVersion Date Revision1.0 000218 Original Release

Pagina 46

AccessoriesB-2B.1B.1 UDAS Termination PanelsUDAS Termination PanelsThe Termination Panel plugs directly into the I/O connectors of the UDAS unit, prov

Pagina 47

B-3FIGURE B.2, PCI-20429T-1 Multifunction Termination Panel shows the signals assigned to each terminal block on this panel.FIGURE B.2 PCI-20429T-1 Mu

Pagina 48

AccessoriesB-4B.1.2B.1.2 UDAS Termination and Signal Conditioning Options UDAS Termination and Signal Conditioning Options FIGURE B.3, UDAS-1001E Term

Pagina 49 - AccessoriesAccessories

B-5FIGURE B.4 UDAS-1001E Termination Options (continued)PCI-20324T-1 Opto-Isolated Digital I/O Panel 8 Channels for use with 1100 Series PCI-20326T-1

Pagina 50 - Accessories

AccessoriesB-6FIGURE B.5 Diagram of the PCI-20430A-1’s connectors and cablesP2P3+ 5 VoltsPort 0: Bit 0 or Counter InDigital GroundPort 1: Bit 0No Conn

Pagina 51

B-7B.1.3B.1.3 Isolator BlocksIsolator BlocksPCI-1100 Series Digital Opto-Isolator BlocksPCI-1100 Series Digital Opto-Isolator BlocksThe PCI-1100 Serie

Pagina 52

C-1Appendix CAppendix C Analog Input and Output Analog Input and Output Calibration ProceduresCalibration ProceduresC.1C.1 UDAS Analog Input Circuit C

Pagina 53 - I/O Connectors

Analog Input and Output Calibration ProceduresC-24. Connect the UDAS to the host computer and allow at least a 15 minute warm-up period before perform

Pagina 54

C-3C.1.2C.1.2 Analog Input Calibration ProcedureAnalog Input Calibration ProcedureCalibrating the analog input section requires adjusting the input of

Pagina 55

Analog Input and Output Calibration ProceduresC-4.7. Set the calibrated input to 9.5068 V.8. Adjust Pot R46 for 9.5068 V.9. Repeat the procedure and v

Pagina 56 - ProceduresProcedures

iChapter 1 UDAS-1001E Series Data Acquisition Systems1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 57 - 00 00 0 0

C-59. Repeat the procedure and verify the adjustments are correct.Adjust Offset and Gain for 0 to 10 Volt RangeAdjust Offset and Gain for 0 to 10 Volt

Pagina 58

Analog Input and Output Calibration ProceduresC-6C.2C.2 UDAS Analog Output Circuit Calibration UDAS Analog Output Circuit Calibration ProcedureProcedu

Pagina 59

C-7Analog Output Channel 0 Offset and GainAnalog Output Channel 0 Offset and Gain1. Verify the termination panel is attached to a 50-pin connector of

Pagina 60

Index-1Aanalog input 1-3, 1-4paced 3-11test 3-5analog output 1-3, 1-8test 3-6auto zero mode 1-8Bbias resistors 2-4block diagram, UDAS system 1-2Ccable

Pagina 61 - ProcedureProcedure

IndexIndex-2LLabView 5.0 support library 1-3libraryLabVIEW Support 1-3U-Link software for Windows 98 1-3Visual Designer Support 1-3library, software s

Pagina 62

Index-3initial connection 2-5models 1-1serial number 3-3UDAS-1001E-1 1-1UDAS-1001E-1G 1-1UDAS-1001E-2 1-1UDAS-1001E-2G 1-1UDAS-1001E-3 1-1UDAS-1001E-4

Pagina 63

Table of Contentsii3.4.1.1 Verify Lower Level Drivers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13Appendix A SpecificationsA.1

Pagina 64

1-1Chapter 1Chapter 1 UDAS-1001E Series Data UDAS-1001E Series Data Acquisition SystemsAcquisition Systems1.11.1 IntroductionIntroductionThe UDAS-1001

Pagina 65

UDAS-1001E Series Data Acquisition Systems1-21.1.11.1.1 FeaturesFeaturesFeatures and basic specifications of the UDAS units are listed below. Specific